A silastic positioner for a modiolus-hugging position of intracochlear electrodes: electrophysiologic effects
- PMID: 10733186
- DOI: 10.1016/s0196-0709(00)80011-3
A silastic positioner for a modiolus-hugging position of intracochlear electrodes: electrophysiologic effects
Abstract
Hypothesis: It was postulated that an electrode array that achieved a close modiolar proximity would result in reduced threshold levels and amplitude slopes, as measured with electrically evoked auditory brainstem responses (EABRs).
Background: Quality and quantity of auditory information transmitted by a cochlear implant to patients with sensorineural hearing loss depend on spatial and temporal resolution achieved by the electrical intracochlear stimulation.
Methods: To improve spatial resolution, a new electrode system was developed by Advanced Bionics Corp., with the intention of obtaining greater modiolar proximity. The implant version specified for animal experiments consists of a straight electrode array of seven embedded platinum discs and a so-called Silastic-positioner. The Silastic positioner is shaped to follow the dimensions of the scala tympani with a concave (triangular) inner side, which fits the form of the electrode array. The aim of the study was to evaluate the influence of a modiolus-hugging electrode position in contrast to a conventional electrode position on EABR in short-term animal experiments. Short-term electrophysiologic studies were performed on six adult cats. After local intracochlear application of neomycin solution (50 mg/mL), electrodes were inserted into the scala tympani. Electrically evoked auditory brainstem response threshold levels and EABR amplitude slopes were systematically investigated with and without the positioner.
Results: Electrically evoked auditory brainstem response measurements revealed a distinct apicobasal threshold shift, with increasing thresholds toward the basal end of the electrode. After insertion of the positioner, this shift diminished or was inverted and EABR thresholds and amplitude slopes were reduced significantly.
Conclusions: Threshold and amplitude slope data emphasize the functional benefit of the positioner system, especially for the stimulation of electrodes in the more basal channels.
Similar articles
-
An overview of cochlear implant electrode array designs.Hear Res. 2017 Dec;356:93-103. doi: 10.1016/j.heares.2017.10.005. Epub 2017 Oct 18. Hear Res. 2017. PMID: 29102129 Review.
-
Electrically evoked compound action potential amplitude growth functions and HiResolution programming levels in pediatric CII implant subjects.Ear Hear. 2004 Dec;25(6):528-38. doi: 10.1097/00003446-200412000-00002. Ear Hear. 2004. PMID: 15604914
-
Electrically evoked auditory brain stem responses for lateral and medial placement of the Clarion HiFocus electrode.Ear Hear. 2003 Apr;24(2):184-90. doi: 10.1097/01.AUD.0000061230.58992.9A. Ear Hear. 2003. PMID: 12677114
-
Chronic electrical stimulation of the auditory nerve at high stimulus rates: a physiological and histopathological study.Hear Res. 1997 Mar;105(1-2):1-29. doi: 10.1016/s0378-5955(96)00193-1. Hear Res. 1997. PMID: 9083801
-
Electrical stimulation of the auditory nerve: the effect of electrode position on neural excitation.Hear Res. 1993 Mar;66(1):108-20. doi: 10.1016/0378-5955(93)90265-3. Hear Res. 1993. PMID: 8473242
Cited by
-
Silastic Electrode Positioner Extrusion as a Late Complication of Cochlear Implantation Surgery.J Int Adv Otol. 2022 Jul;18(4):371-373. doi: 10.5152/iao.2022.21391. J Int Adv Otol. 2022. PMID: 35894535 Free PMC article.
-
Multichannel optogenetic stimulation of the auditory pathway using microfabricated LED cochlear implants in rodents.Sci Transl Med. 2020 Jul 22;12(553):eabb8086. doi: 10.1126/scitranslmed.abb8086. Sci Transl Med. 2020. PMID: 32718992 Free PMC article.
-
Electrophysiological Estimates of the Electrode-Neuron Interface Differ Between Younger and Older Listeners With Cochlear Implants.Ear Hear. 2020 Jul/Aug;41(4):948-960. doi: 10.1097/AUD.0000000000000827. Ear Hear. 2020. PMID: 32032228 Free PMC article.
-
Examination and Comparison of Electrically Evoked Compound Action Potentials and Electrically Evoked Auditory Brainstem Response Results of Children with Cochlear Implantation without Inner Ear Anomaly.Turk Arch Otorhinolaryngol. 2019 Jun;57(2):81-85. doi: 10.5152/tao.2019.4130. Epub 2019 Jun 27. Turk Arch Otorhinolaryngol. 2019. PMID: 31360925 Free PMC article.
-
Individual Optimization of the Insertion of a Preformed Cochlear Implant Electrode Array.Int J Otolaryngol. 2015;2015:724703. doi: 10.1155/2015/724703. Epub 2015 Sep 10. Int J Otolaryngol. 2015. PMID: 26448764 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous