CT-derived estimation of cochlear morphology and electrode array position in relation to word recognition in Nucleus-22 recipients
- PMID: 12382107
- PMCID: PMC3202410
- DOI: 10.1007/s101620020013
CT-derived estimation of cochlear morphology and electrode array position in relation to word recognition in Nucleus-22 recipients
Abstract
This study extended the findings of Ketten et al. [Ann. Otol. Rhinol. Laryngol. Suppl. 175:1-16 (1998)] by estimating the three-dimensional (3D) cochlear lengths, electrode array intracochlear insertion depths, and characteristic frequency ranges for 13 more Nucleus-22 implant recipients based on in vivo computed tomography (CT) scans. Array insertion depths were correlated with NU-6 word scores (obtained one year after SPEAK strategy use) by these patients and the 13 who used the SPEAK strategy from the Ketten et al. study. For these 26 patients, the range of cochlear lengths was 29.1-37.4 mm. Array insertion depth range was 11.9-25.9 mm, and array insertion depth estimated from the surgeon's report was 1.14 mm longer than CT-based estimates. Given the assumption that the human hearing range is fixed (20-20,000 Hz) regardless of cochlear length, characteristic frequencies at the most apical electrode (estimated with Greenwood's equation [Greenwood DD (1990) A cochlear frequency--position function of several species--29 years later. J Acoust. Soc. Am. 33: 1344-1356] and a patient-specific constant as) ranged from 308 to 3674 Hz. Patients' NU-6 word scores were significantly correlated with insertion depth as a percentage of total cochlear length (R = 0.452; r2 = 0.204; p = 0.020), suggesting that part of the variability in word recognition across implant recipients can be accounted for by the position of the electrode array in the cochlea. However, NU-6 scores ranged from 4% to 81% correct for patients with array insertion depths between 47% and 68% of total cochlear length. Lower scores appeared related to low spiral ganglion cell survival (e.g., lues), aberrant current paths that produced facial nerve stimulation by apical electrodes (i.e., otosclerosis), central auditory processing difficulty, below-average verbal abilities, and early Alzheimer's disease. Higher scores appeared related to patients' high-average to above-average verbal abilities. Because most patients' scores increased with SPEAK use, it is hypothesized that they accommodated to the shift in frequency of incoming sound to a higher pitch percept with the implant than would normally be perceived acoustically.
Similar articles
-
Optimal electrode length to match patient specific cochlear anatomy.Eur Ann Otorhinolaryngol Head Neck Dis. 2016 Jun;133 Suppl 1:S68-71. doi: 10.1016/j.anorl.2016.05.001. Epub 2016 May 27. Eur Ann Otorhinolaryngol Head Neck Dis. 2016. PMID: 27246743 Review.
-
Intracochlear Position of Cochlear Implants Determined Using CT Scanning versus Fitting Levels: Higher Threshold Levels at Basal Turn.Audiol Neurootol. 2016;21(1):54-67. doi: 10.1159/000442513. Epub 2016 Feb 18. Audiol Neurootol. 2016. PMID: 26891130
-
Factors affecting open-set word recognition in adults with cochlear implants.Ear Hear. 2013 May-Jun;34(3):342-60. doi: 10.1097/AUD.0b013e3182741aa7. Ear Hear. 2013. PMID: 23348845 Free PMC article.
-
Potential benefits from deeply inserted cochlear implant electrodes.Ear Hear. 2011 Jul-Aug;32(4):411-27. doi: 10.1097/AUD.0b013e3182064bda. Ear Hear. 2011. PMID: 21248642 Review.
-
In vivo measures of cochlear length and insertion depth of nucleus cochlear implant electrode arrays.Ann Otol Rhinol Laryngol Suppl. 1998 Nov;175:1-16. Ann Otol Rhinol Laryngol Suppl. 1998. PMID: 9826942
Cited by
-
Effect of Cochlear Implant Electrode Insertion Depth on Speech Perception Outcomes: A Systematic Review.Otol Neurotol Open. 2023 Dec 14;3(4):e045. doi: 10.1097/ONO.0000000000000045. eCollection 2023 Dec. Otol Neurotol Open. 2023. PMID: 38516541 Free PMC article.
-
Development of a feline model for preclinical research of a new translabyrinthine auditory nerve implant.Front Neurosci. 2024 Feb 6;18:1308663. doi: 10.3389/fnins.2024.1308663. eCollection 2024. Front Neurosci. 2024. PMID: 38379760 Free PMC article.
-
Macrophage depletion attenuates degeneration of spiral ganglion neurons in kanamycin-induced unilateral hearing loss model.Sci Rep. 2023 Oct 5;13(1):16741. doi: 10.1038/s41598-023-43927-9. Sci Rep. 2023. PMID: 37798459 Free PMC article.
-
Electrode array positioning after cochlear reimplantation from single manufacturer.Cochlear Implants Int. 2023 Sep;24(5):273-281. doi: 10.1080/14670100.2023.2179756. Epub 2023 Feb 22. Cochlear Implants Int. 2023. PMID: 37489512
-
Comparing linear and non-linear models to estimate the appropriate cochlear implant electrode array length-are current methods precise enough?Eur Arch Otorhinolaryngol. 2024 Jan;281(1):43-49. doi: 10.1007/s00405-023-08064-z. Epub 2023 Jul 19. Eur Arch Otorhinolaryngol. 2024. PMID: 37466660 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Research Materials