Use of polyacrylamide gels in a saline-filled tank to determine the linearity of the Sheffield Mark 1 electrical impedance tomography (EIT) system in measuring impedance disturbances
- PMID: 8087049
- DOI: 10.1088/0967-3334/15/2a/006
Use of polyacrylamide gels in a saline-filled tank to determine the linearity of the Sheffield Mark 1 electrical impedance tomography (EIT) system in measuring impedance disturbances
Abstract
Polyacrylamide gels have been assessed as a means of producing reliable small impedance changes for EIT studies in saline-filled tanks. Test rods could be made in about 10 min from a stock solution of acrylamide monomer and bis-acrylamide, ammonium persulphate and tetramethylethylaminediamine, completed with saline. Impedance increases of about 10%-200% were produced by varying the gel concentration. The method appears to be easy to use, inexpensive and is stable over time. Its use was illustrated by acquiring EIT images with a Sheffield Mark 1 system and a tank filled with 0.2% saline. Polythene or polyacrylamide rods of varying resistivities, 10% of the tank diameter wide, were placed in one of three positions. The magnitude of the impedance increase in EIT images varied linearly with the actual impedance change in the rods for increases up to 20%, but the blurring in images was similar when polythene rods were used. This suggests that the use of highly resistive test objects gives a reasonable indication of the blurring in EIT images acquired with the system used, but measurement of the magnitude of the impedance change is more likely to be accurate for changes of a few tens of per cent.
Similar articles
-
Bioimpedance tomography (electrical impedance tomography).Annu Rev Biomed Eng. 2006;8:63-91. doi: 10.1146/annurev.bioeng.8.061505.095716. Annu Rev Biomed Eng. 2006. PMID: 16834552 Review.
-
Electrical impedance tomography spectroscopy (EITS) for human head imaging.Physiol Meas. 2003 May;24(2):477-89. doi: 10.1088/0967-3334/24/2/358. Physiol Meas. 2003. PMID: 12812431
-
Validation of a 3D reconstruction algorithm for EIT of human brain function in a realistic head-shaped tank.Physiol Meas. 2001 Feb;22(1):177-85. doi: 10.1088/0967-3334/22/1/321. Physiol Meas. 2001. PMID: 11236878
-
Factors affecting electrode-gel-skin interface impedance in electrical impedance tomography.Med Biol Eng Comput. 1996 Nov;34(6):397-408. doi: 10.1007/BF02523842. Med Biol Eng Comput. 1996. PMID: 9039740 Review.
-
Correction of the non-uniform spatial sensitivity of electrical impedance tomography images.Physiol Meas. 1994 May;15 Suppl 2a:A147-52. doi: 10.1088/0967-3334/15/2a/019. Physiol Meas. 1994. PMID: 8087037
Cited by
-
A novel two-dimensional phantom for electrical impedance tomography using 3D printing.Sci Rep. 2024 Jan 24;14(1):2115. doi: 10.1038/s41598-024-52696-y. Sci Rep. 2024. PMID: 38267531 Free PMC article.
-
Full-field, conformal epiretinal electrode array using hydrogel and polymer hybrid technology.Sci Rep. 2023 Apr 28;13(1):6973. doi: 10.1038/s41598-023-32976-9. Sci Rep. 2023. PMID: 37117214 Free PMC article.
-
Fast detection and data compensation for electrodes disconnection in long-term monitoring of dynamic brain electrical impedance tomography.Biomed Eng Online. 2017 Jan 7;16(1):7. doi: 10.1186/s12938-016-0294-7. Biomed Eng Online. 2017. PMID: 28086909 Free PMC article.
-
A versatile high-permittivity phantom for EIT.IEEE Trans Biomed Eng. 2008 Nov;55(11):2601-7. doi: 10.1109/TBME.2008.2001287. IEEE Trans Biomed Eng. 2008. PMID: 18990630 Free PMC article.
-
[Electrical impedance tomography: ready for routine clinical use for mechanically ventilated patients?].Anaesthesist. 2008 Jan;57(1):61-9. doi: 10.1007/s00101-007-1273-y. Anaesthesist. 2008. PMID: 17934702 Review. German.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous