The contour electrode array: safety study and initial patient trials of a new perimodiolar design
- PMID: 11314713
- DOI: 10.1097/00129492-200101000-00007
The contour electrode array: safety study and initial patient trials of a new perimodiolar design
Abstract
Objective: The aim of these studies was to investigate the insertion properties and safety of a new intracochlear perimodiolar electrode array design (Contour).
Background: An electrode array positioned close to the neural elements could be expected to reduce stimulation thresholds and might potentially reduce channel interaction.
Methods: Two sequential studies were conducted. In study 1, the Contour electrode array was inserted in 12 human temporal bones. After cochlear surface preparation, the position of the array was noted and the basilar membrane was examined for insertion damage. On the basis of the outcome of this temporal bone study, study 2 investigated the Contour array, mounted on a Nucleus CI-24 M device and implanted in three adult patients.
Results: Study I showed that in 10 temporal bones, the Contour array was positioned close to the modiolus, and the basilar membrane was intact. In the two remaining bones, the arrays had pierced the basilar membrane and were positioned in the scala vestibuli apical to the penetration. Statistical analysis showed an equivalent probability of insertion-induced damage of the two array designs. In study 2, image analysis indicated that the Contour electrodes were positioned closer to the modiolus than the standard Nucleus straight array. Lower T and C levels, but higher impedance values, were recorded from electrodes close to the modiolus. Initial speech perception data showed that all patients gained useful open-set speech perception, two patients achieving scores of 100% on sentence material 3 months postoperatively.
Conclusions: The temporal bone studies showed the Contour electrode array to be generally positioned closer to the modiolus than the standard Nucleus straight array, and to have an equivalent probability of causing insertion-induced damage.
Similar articles
-
An overview of cochlear implant electrode array designs.Hear Res. 2017 Dec;356:93-103. doi: 10.1016/j.heares.2017.10.005. Epub 2017 Oct 18. Hear Res. 2017. PMID: 29102129 Review.
-
Surgical implications of perimodiolar cochlear implant electrode design: avoiding intracochlear damage and scala vestibuli insertion.Cochlear Implants Int. 2001 Sep;2(2):135-49. doi: 10.1179/cim.2001.2.2.135. Cochlear Implants Int. 2001. PMID: 18792095
-
[Evaluation of the insertion-trauma of the Nucleus Contour Advance electrode-array in a human temporal bone model].Laryngorhinootologie. 2004 Dec;83(12):840-4. doi: 10.1055/s-2004-826067. Laryngorhinootologie. 2004. PMID: 15611904 German.
-
The Nucleus Contour electrode array: a radiological and histological study.Laryngoscope. 2001 Mar;111(3):508-14. doi: 10.1097/00005537-200103000-00023. Laryngoscope. 2001. PMID: 11224784
-
Comparison of electrode position in the human cochlea using various perimodiolar electrode arrays.Am J Otol. 2000 Mar;21(2):205-11. doi: 10.1016/s0196-0709(00)80010-1. Am J Otol. 2000. PMID: 10733185
Cited by
-
Cochlear implant electrode design for safe and effective treatment.Front Neurol. 2024 May 2;15:1348439. doi: 10.3389/fneur.2024.1348439. eCollection 2024. Front Neurol. 2024. PMID: 38756216 Free PMC article. Review.
-
The geometry of photopolymerized topography influences neurite pathfinding by directing growth cone morphology and migration.J Neural Eng. 2024 Apr 4;21(2):026027. doi: 10.1088/1741-2552/ad38dc. J Neural Eng. 2024. PMID: 38547528 Free PMC article.
-
In vitro impact of platinum nanoparticles on inner ear related cell culture models.PLoS One. 2023 Apr 24;18(4):e0284794. doi: 10.1371/journal.pone.0284794. eCollection 2023. PLoS One. 2023. PMID: 37093819 Free PMC article.
-
A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds.Materials (Basel). 2022 Apr 22;15(9):3049. doi: 10.3390/ma15093049. Materials (Basel). 2022. PMID: 35591381 Free PMC article.
-
Dendritic Degeneration of Human Auditory Nerve Fibers and Its Impact on the Spiking Pattern Under Regular Conditions and During Cochlear Implant Stimulation.Front Neurosci. 2020 Nov 19;14:599868. doi: 10.3389/fnins.2020.599868. eCollection 2020. Front Neurosci. 2020. PMID: 33328872 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical