Perceptual efficacy of electrical stimulation of human retina with a microelectrode array during short-term surgical trials
- PMID: 14638739
- DOI: 10.1167/iovs.02-0817
Perceptual efficacy of electrical stimulation of human retina with a microelectrode array during short-term surgical trials
Abstract
Purpose: This work is part of a feasibility assessment of a retinal prosthesis as a means to restore vision to patients with blindness caused by retinitis pigmentosa. The primary goal was to assess the concordance of the form of induced perception and the pattern of electrical stimulation of the retina, and the reproducibility of the responses.
Methods: Five volunteers with severe retinitis pigmentosa and one with normal vision were studied. A companion paper in this issue provides details on demographics, visual function, surgical methods, general stimulation strategy, and data analysis. Volunteers were awake during surgery while a 10-microm-thick, microfabricated electrode array was placed on the retina. The array was connected to extraocular current sources that delivered charges to 50-, 100-, and 400-microm-diameter electrodes. Negative control trials were randomly included. Perceptual quality was judged by the similarity between the form of stimulation and perception (i.e., accuracy) and the reproducibility of responses.
Results: Only 1 of 40 control tests yielded a false-positive result. On average, volunteers 3, 5, and 6 reported percepts that matched the stimulation pattern 48% and 32% of the time for single- and multiple-electrode trials, respectively. Two-point discrimination in the best cases may have been achieved in two blind subjects using (center-to-center) electrode separation of 600 and 1960 microm. Reproducibility was achieved 66% of the time in the blind subjects. By comparison, in the normal-sighted subject, perceptual form was reported accurately 57% of the time, with 82% reproducibility, and two-point discrimination may have been achieved in one trial with 620-microm electrode spacing and in two trials each with 1860- and 2480-microm electrode spacing. In subjects 5 and 6, perceptual size was inconsistently related to the charge, although relatively large differences in charge (median: 0.55 microcoulombs [microC]) between two trials produced differently sized percepts. Longer stimuli did not produce rounder percepts.
Conclusions: Single percepts induced by single-electrode stimulation were relatively small, but the form of percepts, especially after multielectrode stimulation, often did not match the stimulation pattern, even in a normal-sighted volunteer. Reproducible percepts were more easily generated than those that matched the stimulation pattern.
Similar articles
-
Electrical Stimulation of the Retina to Produce Artificial Vision.Annu Rev Vis Sci. 2016 Oct 14;2:273-294. doi: 10.1146/annurev-vision-111815-114425. Annu Rev Vis Sci. 2016. PMID: 28532361 Review.
-
[Retinal and Cortical Activation by Electrical Stimulation with Retina Implants].Klin Monbl Augenheilkd. 2016 Nov;233(11):1213-1221. doi: 10.1055/s-0042-116321. Epub 2016 Nov 17. Klin Monbl Augenheilkd. 2016. PMID: 27855445 Review. German.
-
Stimulation with a wireless intraocular epiretinal implant elicits visual percepts in blind humans.Invest Ophthalmol Vis Sci. 2011 Jan 21;52(1):449-55. doi: 10.1167/iovs.09-4410. Invest Ophthalmol Vis Sci. 2011. PMID: 20861492 Clinical Trial.
-
Spatiotemporal interactions in retinal prosthesis subjects.Invest Ophthalmol Vis Sci. 2010 Feb;51(2):1223-33. doi: 10.1167/iovs.09-3746. Epub 2009 Sep 9. Invest Ophthalmol Vis Sci. 2010. PMID: 19741248 Free PMC article.
-
Methods and perceptual thresholds for short-term electrical stimulation of human retina with microelectrode arrays.Invest Ophthalmol Vis Sci. 2003 Dec;44(12):5355-61. doi: 10.1167/iovs.02-0819. Invest Ophthalmol Vis Sci. 2003. PMID: 14638738
Cited by
-
Axonal stimulation affects the linear summation of single-point perception in three Argus II users.J Neural Eng. 2024 Apr 8;21(2):026031. doi: 10.1088/1741-2552/ad31c4. J Neural Eng. 2024. PMID: 38457841 Free PMC article.
-
Seeing the Future: A Review of Ocular Therapy.Bioengineering (Basel). 2024 Feb 13;11(2):179. doi: 10.3390/bioengineering11020179. Bioengineering (Basel). 2024. PMID: 38391665 Free PMC article. Review.
-
Hybrid Neural Autoencoders for Stimulus Encoding in Visual and Other Sensory Neuroprostheses.Adv Neural Inf Process Syst. 2022 Dec;35:22671-22685. Adv Neural Inf Process Syst. 2022. PMID: 37719469 Free PMC article.
-
Retinal ganglion cells undergo cell type-specific functional changes in a computational model of cone-mediated retinal degeneration.Front Neurosci. 2023 May 18;17:1147729. doi: 10.3389/fnins.2023.1147729. eCollection 2023. Front Neurosci. 2023. PMID: 37274203 Free PMC article.
-
High amplitude pulses on the same charge condition efficiently elicit bipolar cell-mediated retinal ganglion cell responses in the degenerate retina.Biomed Eng Lett. 2023 Jan 31;13(2):129-140. doi: 10.1007/s13534-023-00260-4. eCollection 2023 May. Biomed Eng Lett. 2023. PMID: 37124107 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources