Multicenter Study
. 2019 Mar;17(3):525-531.
doi: 10.1111/jth.14383. Epub 2019 Feb 20.

An international external quality assessment for laboratory diagnosis of heparin-induced thrombocytopenia

Affiliations
Free article
Multicenter Study

An international external quality assessment for laboratory diagnosis of heparin-induced thrombocytopenia

Julia J M Eekels et al. J Thromb Haemost. 2019 Mar.
Free article

Abstract

Essentials A pilot study for External Quality Assessment for testing of HIT is described. The qualitative accordance for the PF4/heparin IgG test was 97.6%. The qualitative accordance for functional HIT tests was considerably lower. External Quality Assessment for functional HIT tests is required. SUMMARY: Objective Heparin-induced thrombocytopenia (HIT) is a potentially life-threatening complication of heparin exposure. Diagnosis is most reliable using a combination of an enzyme immunoassay (EIA) that detects antibodies against platelet factor 4 (PF4)/heparin complexes ("antigen" assay) and a "functional" assay that detects platelet-activating properties of the pathogenic HIT antibodies. No External Quality Assessment (EQA) is available for a combination of the tests. Here we report on the results of the first international EQA. Methods The pilot EQA was organized by the Department of Transfusion Medicine, Universitätsmedizin Greifswald, Germany. Six serum samples of patients, which were referred to Greifswald for HIT diagnosis, and one negative control sample were distributed to seven participants in Germany, Canada, and Singapore. Participants were asked to report the optical density (OD) values of their local EIA test for IgG-specific antibodies against the PF4/heparin complexes and the results for a functional assay (HIPA or SRA). Consensus was defined as a minimum 70% agreement, i.e., agreement among at least five of the seven participating laboratories. Results and conclusion Six out of seven participants reported results for EIA, with a high quantitative accordance (97.6%). For the functional assay, consensus was reached for all samples except the negative control, for which some participants reported nonspecific reactivity. All HIT-negative samples were correctly diagnosed by all participants; for HIT-positive samples, consensus of 70% was reached. Although the limited availability of sample material is an obstacle to overcome, an EQA combining both EIA and functional testing is feasible.

Keywords: blood platelets; heparin; laboratory proficiency testing; platelet factor 4; thrombocytopenia.